当前位置: 首页 > 商品列表

Lumigen HyPerBlu provides sensitive chemiluminescent detection of hydrogen peroxide

Abstract

Lumigen HyPerBlu reagent is a one solution chemical substrate based on dioxetane chemiluminescent technology. With HyPerBlu reagent, a direct reaction with hydrogen peroxide produces an intense sustained luminescent signal making it an excellent choice for examination of oxidase activity. The signal intensity is sufficient for measurement on PMT based plate reading luminometers or CCD imaging systems. HyPerBlu technology detection is suitable for automation and miniaturization to 384-well or 1536-well format. HyPerBlu reagent’s wide tolerance for common assay additives (e.g., DMSO), lack of intermediaries, chemical specificity, one-component structure, and excellent stability lead to assay simplicity and excellent robustness.

High-throughput screening (HTS) assays present a variety of unique challenges for assay technologies. Naturally fluorescing compounds, biological intermediaries, and target labels can decrease the robustness of an assay. Lumigen HyPerBlu technology avoids these challenges utilizing a technology based on the specific reaction of a dioxetaneboronic acid with hydrogen peroxide. The resulting production of light is based on the half-life of the intermediate compound. Traditional assay methodologies for the measurement of hydrogen peroxide use horseradish peroxidase (HRP) to catalyze oxidation of colorimetric or fluorescent compounds. These drawbacks limit their stability and reproducibility.

Materials

  • PMT luminometer (Labsystems)
  • 384-well white microtiter plates (Generic)
  • Urea peroxide standard
  • Oxidase enzyme
  • Lumigen HyPerBlu Chemiluminescent Reagent

Protocols

  1. Prepare well with oxidase and analog as appropriate for chosen study.
  2. Add 5 µL standard or sample to each well.
  3. Incubate per oxidase characteristics.
  4. Add 5 µL Lumigen HyPerBlu Chemiluminescent Reagent.
  5. Incubate 20–30 min at room temperature.
  6. Detect with 1-s integration per well for PMT luminometer systems, or 5 s per plate for CCD imaging systems.

Data Analysis

Lower detection limits (LDL) and dynamic range for Lumigen HyPerBlu was determined by preparing a dilution series of urea peroxide standards. The signal obtained was plotted against the standard concentrations tested in the assay. The LDL was determined by using the background mean + 3 × sd.

Glucose Oxidase Test Assay

Glucose was oxidized with glucose oxidase in aqueous solution and the peroxide generated was assayed with Lumigen HyPerBlu (Figure 2). Solutions containing 1.5 × 10-13 to 4.5 × 10-16 moles of glucose oxidase were analyzed in triplicate, in a one-step format. Wells were incubated at 25ºC with a solution of 0.1 M glucose. Chemiluminescence intensity was measured at 22 min.

The glucose oxidase assay shows a linear response over the tested range (Figure 3). This same method could be used in a study of enzymatic response to various conditions or in a detection test for glucose itself.

Results and discussion

As shown in Figure 1, an LDL of 5 × 10-13 moles of Hydrogen Peroxide was achieved. The wide linear dynamic range of 5 logs gives this assay great flexibility to measure a variety of processes with minimal process scaling.

Stability of performance of Lumigen HyPerBlu when stored at 2–8ºC was shown to have a variation of <5% in one year. When Lumigen HyPerBlu is kept at 23ºC, the product stability shows a variation of <3% in 31 days. This ease of use can be compared favorably with traditional peroxide detection technologies that require biological detection components.

================  蚂蚁淘在线  ================

免责声明:本文仅代表作者个人观点,与本网无关。其创作性以及文中陈述文字和内容未经本站证实,对本文以及其中全部或者部分内容、文字的真实性、完整性、及时性本站不做任何保证或承诺,请读者仅作参考,并请自行核实相关内容

版权声明:未经蚂蚁淘在线授权不得转载、摘编或利用其他方式使用上述作品。已经经本网授权使用作品的,应该授权范围内使用,并注明“来源:蚂蚁淘在线”。违反上述声明者,本网将追究其相关法律责任。