GENERALINFORMATION
ThePointe180IIisasmallbenchtopanalyzerdesignedtoreadandcalculatetheresultsofin-vitroclinicaldiagnosticassaysonhumanserum,plasma,orurineaswellasanyotherapplicationrequiringabsorbanceorconcentrationreADIngs.Theanalyzerisprovidedwith43pre-programmedtestparametersandiscapableofstoringupto120assaysinmemoryforlaterrecall.Theanalyzerisequippedwithabuiltinvacuumpumpandexternalwastebottlewithlevelsensing.Theanalyzerissuppliedwitharemovableflowcelltoallowforuseof1cmsquareaswellas12mmroundcuvettes.Theanalyzerisdesignedwithmanyfeaturestominimizeoperatorerror,suchasstablefactorycalibration,pre-programmedcalculations,visualandaudIBLefeedback,flagsanderrormessages.Inaddition,itwillstoreupto512patientresults,512controlresults,20patientsonaworklistand15testsperpatient.*
*Notavailableforclinicalhumanbloodtestinginthe US
ORDERING INFORMATION
Catalog#: | Product: |
90.P3304-180 | Pointe180II |
FEATURES
BichromaticPhotometer:
Differentialwavelengthstocorrectimperfectionsintesttubesorsquarecuvettesandtohelpremoveinterferingsubstances.Sixstandardfilterswithtwooptionalfilters(340-700nm)
Temperaturecontrolledcuvette:
Constant37°Ccuvetteandcuvettewellwhenoptionaltubeorsquarecuvettesmodesareused.
InternalTemperaturecontrol:
Theanalyzermonitorstheincubatorandcuvettetemperatures.Thetemperaturesareautomaticallycorrectedandtheactualtemperatureindegrees(C)isdisplayedonthescreen.
LampSaverFeature:
Theinstrumentautomaticallyshutsoffthelampiftheinstrumenthasnotbeenusedfor15minutes.Thisfeaturecanpreservethelampforuptotwoyears,dependingontheactualusage.
QCTrackingFeature:
TheinstrumentsoftwarecanstorecontrolvaluesandcreateLevy-Jenningsplots.
WorklistandPatientreports:
Theinstrumentcancreate,storedandprintworklistandpatientreportstoon-boardthermalprintertoexternalprinterandPC.
DifferentCuvettemodes:
Theanalyzerflowcellisremovabletoallowuseofround12mmglasstubeor1cmsquarecuvettes.
Flip-updisplay:
Large240x128pixelgraphicaldisplay.
PreprogrammedChemistryProcedures:
OptimizedparametersforPointeScientificreagentsarepreprogrammed.120openchannels.
KineticPlots:
Showskineticplotsinrealtime.
MaintenanceFree:
Theanalyzerdoesnotrequireanyscheduledmaintenanceprocedures.
SPECIFICATIONS
SpectrophotometerType: | Filter |
OpticalConfiguration: | Singlebeamwithcontinuouslyrotatingfilterwheel.Monochromaticorbichromaticreading8filterpositions. |
UsableSpectralRange: | 330to770nm |
SystemProcedures: | OpenandStoredMenu. |
CalculatingModes: | -Absorbance |
| -SingleStandard |
| DifferentialSamples |
| -FactorMode |
| DifferentialSamples |
| -Multi-StandardMode (upto7standards) |
| DifferentialSamples |
| -Multi-Standard% ABS(upto7standards) |
| DifferentialSamples |
| -KineticMode(Consecutively,orSimultaneously(Batch)) |
| ByFactororStandard |
| -FixedTimeKinetic |
| ByFactororStandard |
| -IndexMode |
Channels: | 120open |
SourceofRadiation: | TungstenHalogen,10watt,withautomaticlampsaver. |
SelectionifWavelength: | ByFilter. |
FilterType: | 4-cavityinterference,long-lifeionbeam-assisteddeposition. |
WavelengthAccuracy: | ±3nm |
FilterLocation: | Aftersample(heatabsorbingfilterbeforesample). |
FilterSelection: | Automaticbysoftwareorviakeyboard. |
Wavelengths: | 340,405,505,545,580,630,nmsuppliedstandard,othersoptional. |
HalfBandwidth: | |
1/100bandwidth: | 14nmat340nm. |
FalseRadiantEnergyRatio: | <0.001at340and405nm. |
Cuvette: | 1cmsquare,12mmround,flowthrough. |
TypeSupplied: | Flowthrough. |
Material: | 316stainless,borosilicatewindows. |
Geometry: | Cylindrical,2.3mmdiax5mm+/-0.05mm. |
IlluminatedVolume: | 21ul |
MinimumReadVolume: | 250ul |
Aspiration/Purge: | Vacuumpumpat18cmofHg. |
Valve: | Siliconepinchtype |
CuvetteHolder: | Thermostaticallycontrolledcompartmentat37°C. |
Detector: | Gallium-Arsenide-Phosphidephotodiode. |
DisplayType: | 240x128graphicLCDw/backlight |
Absorbance: | -0.5to3.5(flowthroughmode) |
Concentration: | Maximum999,999 |
KineticResults: | Abs/minwithresolutionof0.0002A/min |
ZeroCompensation: | Automatic |
Range: | -0.5to2.5absorbance |
Parallel: | Centronics/IBM-PCcompatible |
Serial: | RS-232at9600baud,8data,1stop,noparityBi-directional. |
DataInput: | 1)20keykeypad,2)PS2101keyboard(connectoronback) |
SpectrophotometricInaccuracy |
|
FlowThrough: | <0.5%at1absorbance,340/630nmNADHSolution. |
StABIlity: | Betterthan0.003A/hrmonochromaticafterwarmup. |
WarmupTime: | 90secondsphotometric,15minutesfortemperaturecompartment. |
Electronics: | Z180Microprocessor18MHz,128kEPROM,32KBytesNVRAM. |
Power: | Auto-SwitchingPowersupply. |
| Fuses: |
| 2.5A/250Vfast5-20mmglassfuse |
| 2.0A/250VFast5-20mmCeramicFuse |
| (2)6/10250Vslowblow3AGFuses |
DimensionsandWeight: | 40cm(L)x37cm(W)x14cm(H)lidclosed(30cmlidopened),6.4kg |
SpaceRequirements: | 10cmclearanceonallsides. |
EnvironmentalConditions forSafeOperation: | IndoorUseAltitudeupto2000m. |
RecommendedOperatingTemperature: | 15-35°C |
RecommendedOperatingHumidity: | Between10and85%,non-condensing |