REF14782
Automaticchangeoverunit
- GasmonitoringforCO2,N2andmixedgas(max.21 %O2)
- Monitorsthegassupplyandswitchesoverautomaticallytosecondcylinderifsupplyisexhausted
- Ensurescontinuousgassupplywithoutinterruption
- Monitoringofgaspressures
- Opticalandacousticalalarmifpressureistoolowortoohigh
- Graphicdisplaywithtouchscreen
Gasmonitor_english_7_012017.pdf
- Technicaldata
- Options
- BasicSupply
Dimensions | 195x53x181mm(wxdxh) |
Pressurerange | 0.8tomax.1.5bar |
AlarmthresholdandLOWswitchingpoint | P<0.3bar |
AlarmthresholdHIGH | P≥2.0bar |
Maximumpermittedconnectionpressure | P=2.0bar |
Remotealarm | |
Powersupply | 12VbyPlug-inpowersupply100-240VAC |
Interfaces | RS485interfacefordataloggingwithoptionalavailablePCsoftwareLaboDat+ Remotealarm |
REF14943
Supportstandforgasmonitor
|
Gasmonitor
Quickcouplersandfurtherconnectors
Holderforinstallationatthewall
Mainsadapter
Magneticsignsforlabellingofgascylinder