REF14209(withoutO₂control)
REF14452(withO₂control)
Incubatorwithaninnervolumeof16liters
AdvantagesofC16:
- Approvedmedicaldevice
- CO₂sensor:
- DirectHeating:
- Verycompactandspacesavingconstruction
Dualbeaminfraredsensor
Veryprecisemeasurement
Veryshortrecoverytimes
Long-termstable
Independentover-allheatingofinnerwallsanddoor
Homogeneoustemperaturedistributionintheinterior
Veryshortrecoverytimes
C16_english_5_012017.pdf
- Features
- Technicaldata
- Options
Interior | Electro-polishedstainlesssteelwithcovedcorners Shelvesandductworkeasilyremovableforcleaningpurposes |
Temperature | Panelheatedinteriorchamberanddoor |
Humiditysupply | Internalwaterreservoir |
CO₂ | Dualbeaminfraredsensor Measurementsindependentfromtemperatureandhumidity |
Veryshortrecoverytimes | Foralladjustableparametersthroughoptimizedmicroprocessorcontrol |
EasyHandling | Menuguided BacklitLCDdisplay Foilkeypad |
Diagnosticsystem | Opticalandacousticalalarmifsetpointdeviationorsensordefectoccurs Doormonitoring Airmonitoring Hardwareself-test Remotealarm |
Options | O₂control:connectionforN₂ |
Interfaces | Remotealarm(potentialfree) RS485-interfacefordata-loggingwithoptionalPCsoftwareLaboDat+ |
Exterior | Dimensions290x330x480 mm(w x d x h) |
Interior | Volume16 l Electro-polishedstainlesssteel 2shelves205x230 mm(w x d) |
Temperature | Panelheatedinteriorchamberanddoor Range27 °C-42 °Cstarting5 °Caboveambienttemperature StABIlity/uniformity:± 0.1 °C / ± 0.3 °C |
CO₂-control | Dualbeaminfraredsensor Range0–10%CO₂ Stability±0.1%CO₂ |
O₂-control<21%(optional) | Measurementwithgalvanicsensor ConnectionofN₂ Range1%–21%O₂ |
Supply | 110–230VAC,50/60Hz,60W Inletpressureforgas0.8bar Ambienttemperature18 °C - 30 °C |
Classification | SafetyclassI EN61010conform |
OurincubatorC16isavailablewithO2controlasoption(O2<21%) | |
FurtheraccessoriesforincubatorC16 | GasIn-Line-Filter PCsoftwareLaboDat+ |