REF15245
Incubatorwithaninnervolumeof200 liters
AdvantagesofLaboC201:
- 7“colourtouchscreen
- UVdecontamination(option)
- Particlefilter(P3)withinairflow
- Displayofmeasuredvaluesfor2hor24h(T,CO2,rH,O2)
- Cablebushing(Accessport)
- Accesscontrolvianumericalcode(option)
- RS485orethernetinterface
- Activesterilehumidity:
- CO2sensor:
- Directheating:
- Optional:
C201_english_3_082017.pdf
- Features
- Technicaldata
- Options
Interior | Electro-polishedstainlesssteelwithcovedcorners Shelvesandductworkeasilyremovableforcleaningpurpose Cablebushing(Accessport) |
Temperature | Intelligenttemperaturecontrolsystemfordryinnerchamber |
Humiditysupply | Activesterilehumidityviaexternalwatersupplycontainer |
CO2control | Dualbeaminfraredsensor Measurementindependentfromtemperatureandhumidity |
Shortrecoverytimes | Foralladjustableparametersthroughoptimizedmicroprocessorcontrol |
Easyhandling | Menuguided7“colourtouchdisplay Displayofmeasuredvaluesfor2h/24h |
Diagnosticsystem | Opticalandacousticalarmifsetpointdeviationorsensordefectoccurs Doormonitoring Remotealarm |
Options | O2control:connectionforN2possIBLe Subdividedinnerglassdoorwith6smalldoors Accesscontrolvianumericalcode UVdecontamination PCsoftwareLaboDat+ |
Interfaces | Alarmcontact(potentialfree) RS485orEthernetinterface |
Exterior | Dimensions720 x 700 x 870 mm(w x d x h) |
Interior | Volume200 l Electro-polishedstainlesssteel Shelvesheightadjustable |
Temperature | Panelheatedinteriorchamberanddoor Range27 °C - 42 °Cstarting5 °Caboveambienttemperature StABIlity/uniformity± 0.1 °C / ± 0.3 °C Over-temperaturealarm32 °C - 47 °C |
CO2control | Dualbeaminfraredsensor Range0–10 % CO2 Stability± 0.1 % CO2 |
O2control<21 %(optional) | Measurementwithgalvanicsensor ConnectionofN2 Range1 % – 21 % O2 |
Humidity | Activesterilehumiditysupplythroughvaporizingmoduleoperatingat120 °C Measuringrange0–98 % rH/range60 % – 95 % rH |
Accessport | Ø30mm |
Supply | 230 V AC,50/60 Hzor115 V AC,50/60 Hz,250VA Inletpressureforgas0.8 bar Ambienttemperature18 °C - 30 °C |
Classification | SafetyclassI EN61010conform |
Subdividedinnerglassdoor-forevenbetterrecoverytimes Subdividedinnerglassdoorwith6smalldoors Betterrecoverytimesofallparameters Lowergasconsumption
| |
UVdecontamination Individualstartbyuser Duration:180minutes Short-waveUVrADIation(UV-Cradiation,<280nm) Dooropening:UVmodestopsimmediately
| |
Accesscontrolvianumericalcodeontouchscreen
Mechanicalunlockingincaseofpowerfailure(emergencyrelease)
| |
Stackingkit-forsafepositioning
Safe,easyandspace-savingpositioningofthedevicesinthelaboratory
REF10681(supportstand115cm,whitelacqueredsteel,withoutwheels)
REF14411(rollerbase15cm,lockable) | |
OurLaboC201incubatorsareavailablewithO2-control(O2<21 %) | |
Eachofourincubatorspassesanindividualcelltestbeforedelivery. | |
FurtheraccessoriesoftheincubatorLaboC201 | Incubator-Filter-Box GasIn-Line-Filter Accessoriesforactivesterilehumiditysupply PCsoftwareLaboDat+ |